Skip to main content

Table 3 Receiver Operating Characteristics

From: Intraoperative neuromonitoring for function-guided resection differs for supratentorial motor eloquent gliomas and metastases

 

Metastases

Gliomas

Amplitude decline

>50 %

>80 %

>50 %

>80 %

true positive

6.0 %

4.0 %

8.0 %

8.0 %

true negative

70.0 %

89 %

84.0 %

84.0 %

false positive

21.0 %

2.0 %

1.0 %

1.0 %

false negative

4.0 %

6.0 %

6.0 %

6.0 %

negative predictive value

95.0 %

94.0 %

94.0 %

94.0 %

positive predictive value

21.0 %

67.0 %

73.0 %

73.0 %

sensitivity

60.0 %

40.0 %

57.0 %

57.0 %

specificity

77.0 %

98.0 %

97.0 %

97.0 %

  1. Receiver Operating Characteristics (ROC) of MEP monitoring when an amplitude decline of >50 % or >80 % is considered significant depending on tumor type