Ultrasound measurement of optic nerve diameter and optic nerve sheath diameter in healthy Chinese adults

Abstract

Background

Measurement of optic nerve sheath diameter (ONSD) is a fast and non-invasive method in detecting elevated intracranial pressure. However, the reported normal range of ONSD was inconsistent. The objective of the study was to determine the normal range of ONSD in healthy Chinese adults.

Methods

Eyeball transverse diameter (ETD), optic nerve diameter (OND), and ONSD were measured by ultrasound examination in healthy adult volunteers. The OND and ONSD were assessed 3 mm behind the globe. The section showing maximal transverse diameter of the eyeball was frozen and the diameter was measured. Each ETD, OND and ONSD was examined twice and the mean value was calculated.

Results

A total of 519 healthy volunteers were included in the study. The median (interquartile range) of ETD, OND and ONSD were 22.3 (21.6 to 23.1) mm, 3.2 (2.9 to 3.4) mm, and 5.1 (4.7 to 5.4) mm, respectively. The 95 % percentile of ONSD was 5.9 mm. There was no significant difference in ETD, OND or ONSD between male and female, or between left and right eye. ONSD was significantly correlated with OND (r = 0.62, P < 0.001), and the median OND/ONSD ratio (interquartile range) was 0.63 (0.59 to 0.67).

Conclusions

The median and the 95 % percentile of sonographic measurement of ONSD are 5.1 mm and 5.9 mm in healthy Chinese adults. The ONSD is correlated with OND, while independent of gender, age, height, weight and ETD. The median OND/ONSD ratio is 0.63 and this parameter warrants further investigation in patients with brain injury.

Peer Review reports

Background

Intracranial hypertension is a life-threatening syndrome, which is caused by a variety of neurological and non-neurological diseases. Catastrophic deterioration of brain function and death may occur if intracranial hypertension is left undetected and untreated [13]. Monitoring of intracranial pressure (ICP) is recommended in the management of intracranial hypertension [4].

Invasive intracranial devices including intra-ventricular catheters and intra-parenchymal probes are considered the gold standard for ICP measurement [5, 6]. However, invasive ICP monitoring may lead to complications including hemorrhage, infection and catheter blockage or dislodgement. Moreover, it is contraindicated in the situation of coagulopathy or thrombocytopenia [79]. Therefore, non-invasive ICP monitoring has been of a major interest in clinical practice [4].

Although elevated ICP can be detected by computed tomography or magnetic resonance imaging, these techniques are expensive, time-consuming and requiring patient’s transportation. In recent years, measurement of optic nerve sheath diameter (ONSD) by ultrasonography has been developed and suggested as a possible indicator of intracranial hypertension [10]. The optic nerve sheath is a membrane covering the optic nerve behind the eye, and is continuous with the dura mater over the brain. Elevation in ICP can be transmitted through the subarachnoid space, causing extension of the space. The rapid and non-invasive nature of ultrasonography results in its increasingly use for detecting elevated ICP at bedside in the emergency and critical care settings.

However, there has been no consensus on the optimal cut-off value of abnormal ONSD to indicate elevated ICP; and the existing thresholds are only from patients with brain injury. The normal range of ONSD in healthy population is indispensable information to interpret the measurement of ONSD as a marker of intracranial hypertension. Moreover, it is still unknown whether there is a difference of ONSD values between Chinese and other races, due to the lack of related studies. We carried out this prospective observational study to determine the normal range of ONSD in healthy Chinese adults.

Method

Healthy volunteers were prospectively included in the study from October 2014 to January 2015. All subjects were aged 18 years or older. Subjects were excluded if they had a history of neurological disorders, hyperthyroidism or diseases of the optic nerve. Written informed consent was obtained from all subjects before enrollment. The study was approved by the local ethics committee of Fujian Provincial Hospital and performed according to the ethical standards of the latest Declaration of Helsinki.

At study entry, the gender, age, height and weight of the subject were collected. The ultrasound examination was conducted in B-mode by using a Philips Envisor C (M2540A) ultrasound system and a 12–3 MHz linear array probe (Philips Medical Systems, Bothell WA, USA), by one of the two sonographers (GS.D. and YC.Z.). Volunteers were examined in supine position with their head elevated to the angle of 20-30°. A thick layer of gel was applied over the closed upper eyelid. The probe was placed on the gel in the temporal area of the eyelid to prevent pressing the eye. The ETD, OND and ONSD were measured. OND and ONSD were assessed 3 mm behind the globe using an electronic caliper along the axis perpendicularly the retina (Fig. 1). In order to gauge the ETD, horizontal sections of eyeball were obtained by scanning from the superior to the inferior side. The section showing the maximal ETD was frozen and the diameter was measured using the electronic caliper. Each ETD, OND and ONSD was examined twice and the mean value was calculated. We also calculated the OND/ONSD ratio.

Fig. 1
figure 1

Transbulbar sonography of the optic nerve and the optic nerve sheath. OND and ONSD were measured 3 mm behind the globe using an electronic caliper along the axis. The interval between “+” marks was OND, and the interval between “x” marks was ONSD

Statistical analysis was performed by using SPSS 19.0 (IBM Corporation, New York, USA). Basic descriptive statistics were calculated (including median and quartiles, mean ± SD, max, min, and 95 % percentile). Comparison of continuous variables was performed by using Mann–Whitney U-test. Kruskal-Wallis test was used to compare the difference of ONSD between subjects with different heights. Spearman rank correlation test was used to determine the correlation between the ONSD and other parameters including age, gender, height, body mass, OND and the ETD. Two-tailed P values of < 0.05 were considered to be statistically significant.

Results

Baseline characteristics

A total of 519 healthy volunteers were enrolled during the study period, consisting of 207 (39.9 %) males and 312 females (60.1 %). Mean (± SD) age, height and weight of the subjects were 46.1 ± 14.2 years, 163.2 ± 7.7 cm and 60.9 ± 11 kg, respectively.

Measurements of ETD, OND and ONSD

All parameters of eye measurement were non-normal distribution (Fig. 2). The median (IQR) of ETD, OND and ONSD were 22.3 (21.6 to 23.1) mm, 3.2 (2.9 to 3.4) mm, and 5.1 (4.7 to 5.4) mm, respectively (Table 1). In addition, the 95 % percentile of ONSD was 5.9 mm.

Fig. 2
figure 2

Histograms of ETD, OND, ONSD and OND/ONSD ratio. All the measurements were non-normal distributed. The ETD and OND/ONSD ratio were positive-skewed distribution while the OND and the ONSD were negative-skewed distribution

Table 1 Basic descriptive statistics of ETD, OND, ONSD and OND/ONSD ratio

There was no significant difference in ETD, OND and ONSD between the male and female, or between the left and right eye (Table 2). ONSD was significantly correlated with OND (r = 0.62, P < 0.001). The median OND/ONSD ratio (IQR) was 0.63 (0.59 to 0.67). Although a statistically significant correlation was found between ONSD and height, the correlation coefficient was sub-optimal (r = 0.063, P = 0.041). Therefore we compared the ONSD between different height and no difference was observed (P = 0.109, Fig. 3). There was no linear relationship of ONSD with ETD (r = 0.044, P = 0.155), with age (r = 0.023, P = 0.467), or with weight (r = −0.005, P = 0.866).

Table 2 Comparisons of measurements between genders and between left/right eyes
Fig. 3
figure 3

Box plot of the ONSD of different heights. Compared by Kruskal-Wallis test, P = 0.109

Discussion

In this large-sample-size study we identified the normal range of ETD, OND, ONSD and OND/ONSD ratio in Chinese adult population. The main findings were: the ONSD were correlated with OND; the median (IQR) ONSD was 5.1 (4.7 to 5.4) mm and the median OND/ONSD ratio (IQR) was 0.63 (0.59 to 0.67).

The measurement of ONSD by ultrasonography is a rapid, noninvasive and repeatable technique, and the result is not affected by change in position [11]. Moreover, previous studies demonstrated a good intra- and inter-observer reliability of the measurement of ONSD by ultrasonography [12, 13]. It has been suggested as an indicator of intracranial hypertension in recent years [10, 1419]. However, there is still no unified standard threshold of enlarged ONSD to determine intracranial hypertension, which varies from 5.0 mm to 5.9 mm [20].

In studies of healthy volunteers, the range of ONSD measured by ultrasonography also varied obviously. Ballantyne et al. reported that the mean ONSDs measured by 3 observers were 3.2, 3.4 and 3.6 mm, respectively [13]. However, after carefully inspecting the method of measurement in this study, we found that the authors actually reported the OND value, rather than the ONSD [13]. In Romagnuolo et al.’s study conducted in the USA, the mean ONSD varied from 4.4 to 4.8 mm in different positions [11]. Maude et al. reported that median ONSD was 4.41 mm with 95 % of subjects in the range of 4.25 to 4.75 mm in Bangladesh [21]. While Bauerle et al. reported that the mean ONSD was 5.4 mm with a range of 4.3-7.6 mm in German [12].

Compared to previous ultrasonographic studies, we found that our data of OND was similar to Ballantyne et al.’s study, while slightly smaller than Bauerle et al.’s study; the ONSD was greater in our study than Maude et al.’s and Romagnuolo et al.’s studies. The study with largest sample size which measured ONSD using ultrasonography was conducted by Maude et al. in Bangladesh (n = 136) [21]. The median (5.1 mm versus 4.41 mm) and range of ONSD (3.5 to 6.4 mm versus 4.2 to 4.8 mm) in our study was higher than Maude et al.’s report, which compromising 12.5 % subjects with age under 16 years old [22]. This might be the possible explanation of the discrepancy in ONSD measurements. In Romagnuolo et al.’s study, because the authors primarily intended to detect difference with positions rather than to establish the normal range of ONSD, they only enrolled 10 subjects [11]. Data obtained from a smaller sample size may lead to greater sampling error. The difference between these two studies may be due to sampling error.

There are also studies using CT or MRI scan, which can provide high spatial resolution, to determine normal range of ONSD. Vaiman et al. reported the mean distal (3 mm behind the globe) ONSD (range) was 4.94 ± 1.51 (3.5-7.5) mm in the right eye and 5.17 ± 1.34 (3.8-7.9) mm in the left eye in CT scans with healthy Israeli volunteers [22]. Geeraerts et al. reported that the ONSD was 5.08 ± 0.52 mm in T2-weighted MRI in British volunteers [23]. The results of these studies were similar to ours. Previous study had shown a good correlation and agreement between ultrasound and MRI measurement of the ONSD 3 mm behind the papilla [24]. However, it is unknown whether it is comparable between CT and ultrasound measurement of ONSD. While Bauerle et al. reported the ONSD was 5.69 ± 0.77 mm in MRI scan in German [24]. Another study conducted by same authors using ultrasonography also exhibited greater ONSD value in German [12]. Therefore we assume that the difference may be caused by difference of ethnicities.

We also compared our data to the reports of elevated ICP patients. The 95 % percentile of ONSD in our healthy volunteer study was 5.9 mm; whereas the lower bound of 95 % confidence interval of ONSD in those patients with elevated ICP (with a ICP value ≥ 20 mmHg) were ≥ 5.9 mm in most of the studies [14, 1719, 25, 26], except Kimberly et al.’s study (approximate 5.1 mm) [16]. Our data suggested that 95 % of normal Chinese adults have an ONSD value ≤ 5.9 mm and it is reasonable to speculate that an ONSD value > 5.9 mm could be considered as abnormal.

We found that OND correlated with ONSD. It is known that the ONSD is strongly correlated with increased ICP while the OND is not [15]. Since the optic nerve is part of the central nervous system, which is surrounded by a subarachnoid space, elevation in ICP can be transmitted through the subarachnoid space and cause extension of the space (e.g. increased ONSD). The change of correlation of OND with ONSD will occur, and therefore the OND/ONSD ratio might be useful in detecting intracranial hypertension. To the best of our knowledge, there is no study report the OND/ONSD ratio. Our data indicate that OND/ONSD ratio had a relatively narrow normal range and maybe a promising surrogate of ONSD. Our research provides reference data for further investigations in patients with elevated ICP.

Our data suggested that the ONSD and the height were statistically correlated. However, a P value < 0.05 only indicates the rejection of the null hypothesis (rho = 0), a correlation coefficient of 0.063, although statistically significant, was too low and should be considered independent in clinical interpretation. Moreover, we stratified subjects based on different height and the subgroup analysis demonstrated no difference of ONSD between different heights.

This study had several limitations. First, a direct measure of intracranial hypertension was not included thus intracranial pressure was presumed to be normal by taking history. Second, two investigators measured the ONSDs in different patients; therefore it was not possible to determine inter-observer variability in this study. Furthermore, this study included only adult volunteers, and thereby it is unable to determine the normal range of ONSD in pediatrics.

Conclusions

The median (IQR) sonographic measurement of ONSD is 5.1 (4.7 to 5.4) mm at 3 mm behind globe and the 95 % percentile of ONSD is 5.9 mm in healthy Chinese adults. The ONSD is correlated with OND, while independent of gender, age, height, weight and ETD. The median OND/ONSD ratio (IQR) was 0.63 (0.59 to 0.67), and this parameter warrants further investigation in population of patients with brain injury.

Abbreviations

ONSD:

Optic nerve sheath diameter

ETD:

Eyeball transverse diameter

OND:

Optic nerve diameter

IQR:

Interquartile range

ICP:

Intracranial pressure

SD:

Standard deviation

References

  1. Marmarou A, Anderson RL, Ward JD, Choi SC, Young HF, Eisenberg HM, et al. Impact of ICP instability and hypotension on outcome in patients with severe head trauma. J Neurosurg. 1991;75(1S):S59–66. Special Supplements.

    Google Scholar 

  2. Stein DM, Hu PF, Brenner M, Sheth KN, Liu KH, Xiong W, et al. Brief episodes of intracranial hypertension and cerebral hypoperfusion are associated with poor functional outcome after severe traumatic brain injury. J Trauma. 2011;71(2):364–73. discussion 73–4.

    Article  PubMed  Google Scholar 

  3. Vik A, Nag T, Fredriksli OA, Skandsen T, Moen KG, Schirmer-Mikalsen K, et al. Relationship of “dose” of intracranial hypertension to outcome in severe traumatic brain injury. J Neurosurg. 2008;109(4):678–84.

    Article  PubMed  Google Scholar 

  4. Hawthorne C, Piper I. Monitoring of intracranial pressure in patients with traumatic brain injury. Front Neurol. 2014;5:121.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Bratton SL, Chestnut RM, Ghajar J, McConnell Hammond FF, Harris OA, Hartl R, et al. Guidelines for the management of severe traumatic brain injury. VII. Intracranial pressure monitoring technology. J Neurotrauma. 2007;24 Suppl 1:S45–54.

    PubMed  Google Scholar 

  6. Andrews PJ, Citerio G, Longhi L, Polderman K, Sahuquillo J, Vajkoczy P. NICEM consensus on neurological monitoring in acute neurological disease. Intensive Care Med. 2008;34(8):1362–70.

    Article  PubMed  Google Scholar 

  7. Chesnut RM, Temkin N, Carney N, Dikmen S, Rondina C, Videtta W, et al. A trial of intracranial-pressure monitoring in traumatic brain injury. N Engl J Med. 2012;367(26):2471–81.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  8. Mattei TA. Intracranial pressure monitoring in severe traumatic brain injury: who is still bold enough to keep sinning against the level I evidence? World Neurosurg. 2013;79(5–6):602–4.

    Article  PubMed  Google Scholar 

  9. Hartl R, Stieg PE. Intracranial pressure is still number 1 despite BEST:TRIP study. World Neurosurg. 2013;79(5–6):599–600.

    Article  PubMed  Google Scholar 

  10. Frattalone AR, Stevens RD. Intracranial pressure and its surrogates. Intensive Care Med. 2011;37(7):1051–3.

    Article  PubMed  Google Scholar 

  11. Romagnuolo L, Tayal V, Tomaszewski C, Saunders T, Norton HJ. Optic nerve sheath diameter does not change with patient position. Am J Emerg Med. 2005;23(5):686–8.

    Article  PubMed  Google Scholar 

  12. Bauerle J, Lochner P, Kaps M, Nedelmann M. Intra- and interobsever reliability of sonographic assessment of the optic nerve sheath diameter in healthy adults. J Neuroimaging. 2012;22(1):42–5.

    Article  PubMed  Google Scholar 

  13. Ballantyne SA, O’Neill G, Hamilton R, Hollman AS. Observer variation in the sonographic measurement of optic nerve sheath diameter in normal adults. Eur J Ultrasound. 2002;15(3):145–9.

    Article  CAS  PubMed  Google Scholar 

  14. Geeraerts T, Launey Y, Martin L, Pottecher J, Vigue B, Duranteau J, et al. Ultrasonography of the optic nerve sheath may be useful for detecting raised intracranial pressure after severe brain injury. Intensive Care Med. 2007;33(10):1704–11.

    Article  PubMed  Google Scholar 

  15. Geeraerts T, Merceron S, Benhamou D, Vigue B, Duranteau J. Non-invasive assessment of intracranial pressure using ocular sonography in neurocritical care patients. Intensive Care Med. 2008;34(11):2062–7.

    Article  PubMed  Google Scholar 

  16. Kimberly HH, Shah S, Marill K, Noble V. Correlation of optic nerve sheath diameter with direct measurement of intracranial pressure. Acad Emerg Med. 2008;15(2):201–4.

    Article  PubMed  Google Scholar 

  17. Moretti R, Pizzi B. Optic nerve ultrasound for detection of intracranial hypertension in intracranial hemorrhage patients: confirmation of previous findings in a different patient population. J Neurosurg Anesthesiol. 2009;21(1):16–20.

    Article  PubMed  Google Scholar 

  18. Moretti R, Pizzi B, Cassini F, Vivaldi N. Reliability of optic nerve ultrasound for the evaluation of patients with spontaneous intracranial hemorrhage. Neurocrit Care. 2009;11(3):406–10.

    Article  PubMed  Google Scholar 

  19. Soldatos T, Karakitsos D, Chatzimichail K, Papathanasiou M, Gouliamos A, Karabinis A. Optic nerve sonography in the diagnostic evaluation of adult brain injury. Crit Care. 2008;12(3):R67.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Dubourg J, Javouhey E, Geeraerts T, Messerer M, Kassai B. Ultrasonography of optic nerve sheath diameter for detection of raised intracranial pressure: a systematic review and meta-analysis. Intensive Care Med. 2011;37(7):1059–68.

    Article  PubMed  Google Scholar 

  21. Maude RR, Hossain MA, Hassan MU, Osbourne S, Sayeed KL, Karim MR, et al. Transorbital sonographic evaluation of normal optic nerve sheath diameter in healthy volunteers in Bangladesh. PLoS ONE. 2013;8(12), e81013.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Vaiman M, Gottlieb P, Bekerman I. Quantitative relations between the eyeball, the optic nerve, and the optic canal important for intracranial pressure monitoring. Head Face Med. 2014;10:32.

    Article  PubMed  PubMed Central  Google Scholar 

  23. Geeraerts T, Newcombe VF, Coles JP, Abate MG, Perkes IE, Hutchinson PJ, et al. Use of T2-weighted magnetic resonance imaging of the optic nerve sheath to detect raised intracranial pressure. Crit Care. 2008;12(5):R114.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Bauerle J, Schuchardt F, Schroeder L, Egger K, Weigel M, Harloff A. Reproducibility and accuracy of optic nerve sheath diameter assessment using ultrasound compared to magnetic resonance imaging. BMC Neurol. 2013;13:187.

    Article  PubMed  PubMed Central  Google Scholar 

  25. Amini A, Kariman H, Arhami Dolatabadi A, Hatamabadi HR, Derakhshanfar H, Mansouri B, et al. Use of the sonographic diameter of optic nerve sheath to estimate intracranial pressure. Am J Emerg Med. 2013;31(1):236–9.

    Article  PubMed  Google Scholar 

  26. Cammarata G, Ristagno G, Cammarata A, Mannanici G, Denaro C, Gullo A. Ocular ultrasound to detect intracranial hypertension in trauma patients. J Trauma. 2011;71(3):779–81.

    Article  PubMed  Google Scholar 

Download references

Acknowledgement

We thank Yi-Yang Chen and Rong Na for language editing of the manuscript.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jian-Xin Zhou.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

HC, RGY and JXZ participated in the design of the study, data analysis and drafted the manuscript. GSD and YCZ participated in the data collection. All authors edited the manuscript and read and approved the final manuscript.

Rights and permissions

Open Access  This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.

The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

To view a copy of this licence, visit https://creativecommons.org/licenses/by/4.0/.

The Creative Commons Public Domain Dedication waiver (https://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Chen, H., Ding, GS., Zhao, YC. et al. Ultrasound measurement of optic nerve diameter and optic nerve sheath diameter in healthy Chinese adults. BMC Neurol 15, 106 (2015). https://0-doi-org.brum.beds.ac.uk/10.1186/s12883-015-0361-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s12883-015-0361-x

Keywords